Products

PC-1000 Online acidimeter

This online PH meter can be widely used in waste water treatment, pure water, circulating water, boiler water and other electronic and electroplating, printing and dyeing, chemical, food, pharmaceutical and other process areas, in large-scale sewage treatment plants, industrial process monitoring and other applications Excellent performance.

Online Consulting

Online Message

Features

  • Features
  • Technical Specifications
  • Standard Configuration
  • Data download

Features

Technical Specifications

Standard Configuration

Data download

-144X144 microcomputer design pH/redox transmitter
-Outdoor waterproof design suitable for wall, panel, round tube installation
-Large LCD screen with operating status and special symbol display
-Backlight sensor, automatic and manual backlight settings
-Provide electrode automatic cleaning device contact, program setting output
-Provides solution grounding to eliminate sample interference
-With automatic correction; manual / automatic temperature compensation function and no cycle correction warning
-Password lock and combination keys to enter the settings, calibration mode design, and enhance security protection
-With RS-485 interface output, can be directly connected with the digital system (PC-1000RS only)
-With two channels of 0/4~20mA output for setting pH/ORP, Temp.(PC-1001 only)

Measuring range

PH :-2.00-16. 00pH ;

ORP :-1999-+ 1999mV

TEMP : -30.0-130

Resolution

PH : 0.0lpH ;

ORP : -lmV

TEMP : 0.1℃

Accuracy

001pH±1Digit ; 0.1%±1Digit ;

 0.2℃±1Digit has the function to modify temperature error

Temperature compensation

NTC30K/PT1K automatic temperature compensation; manual temperature compensation

Corrective mode

Three-point correction

Operating ambient temperature

0 - 50℃

Input resistance

>1012 Ω

Display screen

Backlit large-scale LCM display has back light inductor to make the selection by self-motion or manual operation

Current output I

Isolated type 0/4 — 20mA can set measuring range of relative PH/ORP, and its maximum load is 600Ω

Current output II

Isolated type 0/4 — 20mA can set measuring range of relative temperature, and its maximum load is 600Ω

Control

Contact output: RELAY ON/OFF contact, 240VAC 0.5A MAX

Setting: the independent control point between LO two groups

Cleaning setting

Contact output, ON 0-99 minutes 59 seconds/ OFF 0-999 hours and 59 minutes

Voltage output

DC±12V,1W max

Power supply

100V一240VAC±10%,50/60Hz,4W max

Installing method

wall mounting type/duct line type/panel type

Size of the machine

110mmX95mmX95mm

Size of hollow

90mmX90mm

Weight

0.5kg


Electrodes, manuals, power cords, certificates, and other accessories

RELATED PRODUCTS